Sec. 84.155 Airflow resistance test; Type C supplied-air respirator, requirements.
The resistance to air flowing from the respirator shall not exceed 25 mm. (1 inch) of water-column height when the air flow into the respiratory-inlet covering is 115 liters (4 cubic feet) per minute.